Small/Wide Angle X-ray Scattering (SAXS/WAXS) Ganesha 300XL (SAXSLAB)

The Small/Wide Angle X-Ray Scattering (SAXS/WAXS) Ganesha 300XL (SAXSLAB) offer many features to sample, below is series of information relating to that.

Introduction

YouTube Widget Placeholder

The CNIE at »Ê¼Ò»ªÈË has a unique in-house Small/Wide Angle X-ray Scattering (SAXS/WAXS) instrument calledÌýGanesha 300XL. This state-of-the-art SAXS/WAXS is developed by SAXSLAB forÌýÌýwith full motorization, extensive automation, large sample chamber and versatile sample area to provide a highly customised instrument for:

Ìý

  • Ambient work with up to 30 solid samples in a single run
  • Temperature dependent analysis between -100 and 350°C
  • Solution scattering in refillable capillaries from 4 to 80°C
  • GISAXS/GIWAXS Analysis
  • Time-resolved Micro-fluidics monitoring in airÌý
  • q range 0.0025Ã…-1Ìýto 3 Ã…-1, 2θ range 2° to 60°

SAXS provides essential information on the structure and dynamics of large molecular assemblies (polymers, colloids and porous materials) in various environments. It is a useful complementary technique to study complex systems such as:

  • Heterogeneous catalysts and adsorbents
  • Membranes, thin films and interfaces
  • Polymer processing/design
  • Micro- and nano-fluidic systems
  • Supramolecular organisation in biological systems
  • Self-assembly of mesoscopic metal particles, colloids, liquid crystals and devices
  • distribution, adsorption and arrangement ofÌýnanoparticles, proteins or surface charges

Application

GaneshaÌý300XLÌýcan be used to evaluate multi-scale structuresÌýsuch asÌýatomic structure (0.2 - 1 nm)Ìýand molecular structure (1 nm to 100 nm). The sample-to-detector distance is changeable in less than one minute with fullÌýmotorization. It's suitable to study materials with different forms including solids (powder, paste, thin film), liquids, liquid-crystals,Ìýand gels (with ordered and disordered structures).

Main applications:

1) Nano-particle size, shapes, distributionÌýand polydispersity analyses.

2)ÌýThree-dimensional protein molecule structure analyses (folding and unfolding).

3)ÌýIdentification of molecular assembly or disassembly.

4)ÌýColloids, metals, cement, oil, polymers, plastics, foods and pharmaceuticals.


Sample PreparationÌý

1. Load you sample into a 1.0mm or 1.5mm borosilicate glass capillary

2. Seal the capillary either with a flame, superglue or wax

3. Load the capillary samples into the samples holder, and label the samples with its position in log book, e.g.

0305 - samples A

0306 - samples B

0307 - sample C


Calibration

Silver Behenate (peak at 0.1076 1/Ã…)

Silver BehenateÌýis the standard for SAXS machines., with a peak in the middle of the traditional SAXS region, a strong scattering pattern and a shelf-life of several years (when kept dry and in the dark). Unfortunately Silver Behenate is NOTÌýthe golden standard that one would hope for because:ÌýSilver Behenate varies from supplier to supplier; Silver Behenate degrades over time when subjected to air, humidity and/or lightSilver; Behenate often shows peak splitting at higher orders, with peaks separated by about 1.7%.ÌýNeverheless a Silver Behenate sample, when regularly refreshedÌýworks out as a nice, well accepted standard for many SAXS applications. The SAXS scattering pattern of a standard silver behenate powder sample wrapped in aluminum paper is shown.


Specification

·ÌýÌýÌýÌýÌýÌýÌýHigh Brilliance Microfocus Source

·ÌýÌýÌýÌýÌýÌýÌýMotorized Collimation System: consisting of JJ X-Ray high precision 4-blade slits with single crystal low-scatter blades. The continuously variable aperture sizes allow for an unlimited choice of resolution.

·ÌýÌýÌýÌýÌýÌýÌýFully Evacuated Flight PathÌýfor minimal scattering

·ÌýÌýÌýÌýÌýÌýÌýOn-axis sample Viewer

·ÌýÌýÌýÌýÌýÌýÌýLarge sample areaÌýwith space for potentially complicated sample environments

·ÌýÌýÌýÌýÌýÌýÌýMotorized sample stageÌýwith XY motion (for sample placement and multiple sample mounting) combined with a Θ-stage for standard SAXs and GISAXs

·ÌýÌýÌýÌýÌýÌýÌýPilatus 300K Solid-State Photon-Counting DetectorÌýproviding ultimate noise reduction, high dynamic range, and radiation hardness for long-term maintenance free operation.

·ÌýÌýÌýÌýÌýÌýÌýProprietary Noise-reduction AlgorithmÌýto remove cosmic background

·ÌýÌýÌýÌýÌýÌýÌýMultiple Insertable Beam stopsÌýfor different configurations

·ÌýÌýÌýÌýÌýÌýÌýFull 3-axis Detector MotionÌýof the detector, achieving sample-to-detector distances between 80 mm and 1450 mm, and allowing for measurement with the beam centred on the detector as well as off-cantered

·ÌýÌýÌýÌýÌýÌýÌýReal-time automated Data-Reduction


Scientific Highlights:

  • To be updated